Adding to the Medtronic portfolio of catheter-based lead solutions, the novel OmniaSecure defibrillation lead allows for precise delivery and placement in the right ventricle
Heart Rhythm 2025: Late-breaking clinical study results evaluating the OmniaSecure lead for investigational use in the LBBAP location show high defibrillation success
GALWAY, Ireland and SAN DIEGO, April 25, 2025 /PRNewswire/ -- Medtronic plc (NYSE: MDT), a global leader in healthcare technology, received U.S. Food and Drug Administration (FDA) approval for the OmniaSecure™ defibrillation lead for placement within the right ventricle. The lead, built on the highly reliable SelectSecure™ Model 3830 pacing lead and delivered via catheter, builds on the Medtronic portfolio of lead solutions designed for precise delivery and placement. The lead connects to an implantable defibrillator, and treats potentially life threatening ventricular tachyarrhythmias, ventricular fibrillation (VT/VF) and bradyarrhythmias. As the world's smallest defibrillation lead (4.7 French, or 1.6mm), the OmniaSecure lead represents a meaningful innovation in electrophysiology, and is indicated for stimulation in the right ventricle for adults and adolescent pediatric patients ages 12 and up, including those with smaller anatomies.
Separately, the company is also studying placing the novel, small-diameter OmniaSecure defibrillation lead in the left bundle branch (LBB) area, which has the potential to enable physiologic pacing to more closely mimic the heart's natural conduction system. Investigational outcomes of this study were presented at Heart Rhythm 2025 in San Diego. The results from the study demonstrate high defibrillation success of 100% at implant when the lead is implanted in the LBB area. Globally, the OmniaSecure defibrillation lead is investigational for use in LBB area and requires FDA approval in the future.
Implantable cardioverter-defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) are the standard for preventing sudden cardiac death. The ICD/CRT-D connects to a defibrillation lead (insulated electrical wire) that forms the electrical conduit between the device and the heart. The lead senses the heartbeat, and transmits signals to the implanted device, which then delivers therapy to correct or interrupt abnormally fast rhythms. The lead must flex with millions of heart contractions over a lifetime.
More on The Californer
Existing defibrillation leads are larger in diameter than the OmniaSecure lead. A larger-diameter lead may increase the potential for downstream complications, such as venous occlusion or tricuspid valve regurgitation.
"FDA approval for the OmniaSecure defibrillation lead furthers our ability to offer physicians and patients a transvenous solution designed to be smaller to help minimize complications−including vascular complications and valve interaction−with strong, reliable lead durability. We engineered the OmniaSecure lead based on the trusted SelectSecure Model 3830 pacing lead, which has been the lead of choice for many physicians for more than 20 years," said Alan Cheng, M.D., chief medical officer of the Cardiac Rhythm Management business, which is part of the Cardiovascular Portfolio at Medtronic. "This milestone underscores our commitment to driving clinical innovations that help patients today while paving the way for future innovations that will usher in the next era of electrophysiology."
Previously, researchers presented late breaking data from the global Lead Evaluation for Defibrillation and Reliability (LEADR) Pivotal Trial showing the OmniaSecure defibrillation lead met its primary safety and effectiveness endpoints and exceeded prespecified performance goals when placed within the right ventricle. The results were presented during Heart Rhythm 2024, simultaneously published in Heart Rhythm, and are the basis of FDA approval for the traditional right ventricular lead placement indication.
Late-Breaking LEADR LBBAP Results Presented at Heart Rhythm 2025
Researchers presented late-breaking results at Heart Rhythm 2025 for the LEADR LBBAP (Lead Evaluation for Defibrillation and Reliability in Left Bundle Branch Area Pacing) study that showed the OmniaSecure defibrillation lead demonstrated high defibrillation success when placed in the LBB area for patients indicated for an ICD or CRT-D. Placing the defibrillation lead in the LBB area is being evaluated as an alternative to right ventricular stimulation for sensing, pacing, cardioversion and defibrillation.
Defibrillation testing conducted in 162 patients at device implantation was successful in 100% of cases, with the study meeting the prespecified efficacy goal of 88%. Of the first 193 patients implanted in the study, the OmniaSecure lead was successfully implanted per protocol in 95.8% of the procedures as reported by physician investigators. There were no procedure-related major complications such as early helix or lead fracture, system revision, or death.
More on The Californer
"The left bundle branch area is emerging as an option for more physiologic pacing for patients who receive an ICD or CRT-D to treat dangerous heart rhythms," said Pugazhendhi Vijayaraman, M.D., cardiac electrophysiologist at Geisinger Wyoming Valley Medical Center in Wilkes-Barre, Pa., who presented the data at the meeting. "The option to place a lead in the left bundle branch area may provide for physiologic pacing by engaging the heart's natural conduction system. These positive preliminary results for the LEADR LBBAP study are encouraging and highlight the potential versatility of the OmniaSecure defibrillation lead."
The LEADR LBBAP trial is a global, prospective, non-randomized, multi-center study. The study has enrolled approximately 300 patients at 24 sites in 11 countries in North America, Europe, Asia and Australia. Patients enrolled in the study indicated for an ICD are being followed out to 3 months while patients indicated for CRT-D are being followed out to 6 months post-implant.
About Medtronic
Bold thinking. Bolder actions. We are Medtronic. Medtronic plc, headquartered in Galway, Ireland, is the leading global healthcare technology company that boldly attacks the most challenging health problems facing humanity by searching out and finding solutions. Our Mission — to alleviate pain, restore health, and extend life — unites a global team of 95,000+ passionate people across more than 150 countries. Our technologies and therapies treat 70 health conditions and include cardiac devices, surgical robotics, insulin pumps, surgical tools, patient monitoring systems, and more. Powered by our diverse knowledge, insatiable curiosity, and desire to help all those who need it, we deliver innovative technologies that transform the lives of two people every second, every hour, every day. Expect more from us as we empower insight-driven care, experiences that put people first, and better outcomes for our world. In everything we do, we are engineering the extraordinary. For more information on Medtronic (NYSE: MDT), visit www.Medtronic.com and follow Medtronic on LinkedIn.
Any forward-looking statements are subject to risks and uncertainties such as those described in Medtronic's periodic reports on file with the Securities and Exchange Commission. Actual results may differ materially from anticipated results.
Contacts:
Joey Lomicky
Public Relations
+1-763-526-2494
Ryan Weispfenning
Investor Relations
+1-763-505-4626
SOURCE Medtronic plc
Heart Rhythm 2025: Late-breaking clinical study results evaluating the OmniaSecure lead for investigational use in the LBBAP location show high defibrillation success
GALWAY, Ireland and SAN DIEGO, April 25, 2025 /PRNewswire/ -- Medtronic plc (NYSE: MDT), a global leader in healthcare technology, received U.S. Food and Drug Administration (FDA) approval for the OmniaSecure™ defibrillation lead for placement within the right ventricle. The lead, built on the highly reliable SelectSecure™ Model 3830 pacing lead and delivered via catheter, builds on the Medtronic portfolio of lead solutions designed for precise delivery and placement. The lead connects to an implantable defibrillator, and treats potentially life threatening ventricular tachyarrhythmias, ventricular fibrillation (VT/VF) and bradyarrhythmias. As the world's smallest defibrillation lead (4.7 French, or 1.6mm), the OmniaSecure lead represents a meaningful innovation in electrophysiology, and is indicated for stimulation in the right ventricle for adults and adolescent pediatric patients ages 12 and up, including those with smaller anatomies.
Separately, the company is also studying placing the novel, small-diameter OmniaSecure defibrillation lead in the left bundle branch (LBB) area, which has the potential to enable physiologic pacing to more closely mimic the heart's natural conduction system. Investigational outcomes of this study were presented at Heart Rhythm 2025 in San Diego. The results from the study demonstrate high defibrillation success of 100% at implant when the lead is implanted in the LBB area. Globally, the OmniaSecure defibrillation lead is investigational for use in LBB area and requires FDA approval in the future.
Implantable cardioverter-defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) are the standard for preventing sudden cardiac death. The ICD/CRT-D connects to a defibrillation lead (insulated electrical wire) that forms the electrical conduit between the device and the heart. The lead senses the heartbeat, and transmits signals to the implanted device, which then delivers therapy to correct or interrupt abnormally fast rhythms. The lead must flex with millions of heart contractions over a lifetime.
More on The Californer
- Historic 1947 Trumpy Yacht Mauretania Featured in Wooden Boat Magazine
- California Music Hall of Fame 2025 Induction Ceremony at Warner Bros. Studios November 15th
- Angel Light Academy Honors 8 Community Members Who Demonstrate Excellence in Leadership and Service 18th Annual Above & Beyond Awards November 8
- Culture, Classical Guitar and Comedy... HealthyLife.net Radio adds new programs
- THINKWARE to Showcase Three Next-Generation Dash Cams at SEMA Show
Existing defibrillation leads are larger in diameter than the OmniaSecure lead. A larger-diameter lead may increase the potential for downstream complications, such as venous occlusion or tricuspid valve regurgitation.
"FDA approval for the OmniaSecure defibrillation lead furthers our ability to offer physicians and patients a transvenous solution designed to be smaller to help minimize complications−including vascular complications and valve interaction−with strong, reliable lead durability. We engineered the OmniaSecure lead based on the trusted SelectSecure Model 3830 pacing lead, which has been the lead of choice for many physicians for more than 20 years," said Alan Cheng, M.D., chief medical officer of the Cardiac Rhythm Management business, which is part of the Cardiovascular Portfolio at Medtronic. "This milestone underscores our commitment to driving clinical innovations that help patients today while paving the way for future innovations that will usher in the next era of electrophysiology."
Previously, researchers presented late breaking data from the global Lead Evaluation for Defibrillation and Reliability (LEADR) Pivotal Trial showing the OmniaSecure defibrillation lead met its primary safety and effectiveness endpoints and exceeded prespecified performance goals when placed within the right ventricle. The results were presented during Heart Rhythm 2024, simultaneously published in Heart Rhythm, and are the basis of FDA approval for the traditional right ventricular lead placement indication.
Late-Breaking LEADR LBBAP Results Presented at Heart Rhythm 2025
Researchers presented late-breaking results at Heart Rhythm 2025 for the LEADR LBBAP (Lead Evaluation for Defibrillation and Reliability in Left Bundle Branch Area Pacing) study that showed the OmniaSecure defibrillation lead demonstrated high defibrillation success when placed in the LBB area for patients indicated for an ICD or CRT-D. Placing the defibrillation lead in the LBB area is being evaluated as an alternative to right ventricular stimulation for sensing, pacing, cardioversion and defibrillation.
Defibrillation testing conducted in 162 patients at device implantation was successful in 100% of cases, with the study meeting the prespecified efficacy goal of 88%. Of the first 193 patients implanted in the study, the OmniaSecure lead was successfully implanted per protocol in 95.8% of the procedures as reported by physician investigators. There were no procedure-related major complications such as early helix or lead fracture, system revision, or death.
More on The Californer
- ICYMI: Coreshell Technologies wins $1 million grand prize at Startup World Cup Grand Finale, highlighting California's entrepreneurial leadership
- Small-Firm Leaders Must Prioritize High-Impact Tasks to Prevent Burnout: New Article in AT
- Stockdale Capital Partners Acquires 266-Unit Class-A Scottsdale, AZ Multifamily Community
- Revenue Optics Ignites AI Revolution in Industrial Distribution
- Arnica Unveils "Arnie AI" to Secure the Future of AI-Driven Software Development
"The left bundle branch area is emerging as an option for more physiologic pacing for patients who receive an ICD or CRT-D to treat dangerous heart rhythms," said Pugazhendhi Vijayaraman, M.D., cardiac electrophysiologist at Geisinger Wyoming Valley Medical Center in Wilkes-Barre, Pa., who presented the data at the meeting. "The option to place a lead in the left bundle branch area may provide for physiologic pacing by engaging the heart's natural conduction system. These positive preliminary results for the LEADR LBBAP study are encouraging and highlight the potential versatility of the OmniaSecure defibrillation lead."
The LEADR LBBAP trial is a global, prospective, non-randomized, multi-center study. The study has enrolled approximately 300 patients at 24 sites in 11 countries in North America, Europe, Asia and Australia. Patients enrolled in the study indicated for an ICD are being followed out to 3 months while patients indicated for CRT-D are being followed out to 6 months post-implant.
About Medtronic
Bold thinking. Bolder actions. We are Medtronic. Medtronic plc, headquartered in Galway, Ireland, is the leading global healthcare technology company that boldly attacks the most challenging health problems facing humanity by searching out and finding solutions. Our Mission — to alleviate pain, restore health, and extend life — unites a global team of 95,000+ passionate people across more than 150 countries. Our technologies and therapies treat 70 health conditions and include cardiac devices, surgical robotics, insulin pumps, surgical tools, patient monitoring systems, and more. Powered by our diverse knowledge, insatiable curiosity, and desire to help all those who need it, we deliver innovative technologies that transform the lives of two people every second, every hour, every day. Expect more from us as we empower insight-driven care, experiences that put people first, and better outcomes for our world. In everything we do, we are engineering the extraordinary. For more information on Medtronic (NYSE: MDT), visit www.Medtronic.com and follow Medtronic on LinkedIn.
Any forward-looking statements are subject to risks and uncertainties such as those described in Medtronic's periodic reports on file with the Securities and Exchange Commission. Actual results may differ materially from anticipated results.
Contacts:
Joey Lomicky
Public Relations
+1-763-526-2494
Ryan Weispfenning
Investor Relations
+1-763-505-4626
SOURCE Medtronic plc
Filed Under: Business
0 Comments
Latest on The Californer
- California: Governor Newsom announces appointments 10.31.2025
- Warrior Foundation Freedom Station Starts Give-A-Thon to Fly Our Warriors "Home for the Holidays!"
- Award-Winning Author Zane Carson Carruth Featured in USA Today for Inspiring Mission to Nurture Young Hearts Through Storytelling
- As Trump continues illegally federalizing the National Guard, California doubles down on challenge
- California: Governor Newsom applauds major win for all Americans as judges find Trump Administration illegally withheld SNAP benefits
- Long Beach Public Library Announces 2026 Youth Poet Laureate Program
- Long Beach Parks, Recreation and Marine to Offer Free Youth Fishing Derby November 15
- Parkchester Oral & Maxillofacial Surgery Celebrates 450+ 5-Star Reviews
- The AI CEO Partners with D3 Hockey News to Elevate the Voice of Division III Hockey Nationwide
- Statement from the Campaign of Theodis Daniel, Republican for U.S. Congress (TX-18)
- California: Governor Newsom announces judicial appointments 10.31.2025
- FreeFast.Food Steps Up for 42 Million SNAP/EBT Users with Free Tacos and Burritos Nationwide
- California: NO TREATS, ALL TRICKS: The Trump Administration is killing the economy
- Ascend in Motion Expands Flat-Rate Coverage to Anaheim
- Attorney Credits Launches New CLE Course: "Mastering Reptile Tactics" with Kate Whitlock, Esq
- Divine Punk Announces Happy Christmas, a Holiday Soundscape by Rebecca Noelle
- $430 Million 2026 Revenue Forecast; 26% Organic Growth; $500,000 Stock Dividend Highlight a Powerful AI & Digital Transformation Story: IQSTEL $IQST
- Ascend in Motion Expands Flat-Rate Coverage Across Los Angeles
- Wzzph Deploys 5-Million-TPS Trading Engine with Hot-Cold Wallet Architecture Serving 500,000 Active Users Across Latin America
- Preston Dermatology & Skin Surgery Center and Dr. Sheel Desai Solomon Dominate Raleigh's Best Awards from The News & Observer